Device for punctures of cavity structures in postoperative period in patients with ventral hernias
A.G. IZMAILOV1, S.V. DOBROKVASHIN1, D.E. VOLKOV1, E.E. LUKOYANYCHEV2, V.A. PYRKOV3, R.F. ZAKIROV3, Sh.A. DAVLET—KILDEEV3
¹Kazan State Medical University, 49 Butlerov Str., Kazan, Russian Federation, 420012
2City Clinical Hospital №7 named after E.L. Berezov, 66а Oktyabr’skoy Revolyutsii Str., Nizhniy Novgorod, Russian Federation, 603011
3Hospital for War Veterans, 5 Isayev Str., Kazan, Russian Federation, 420039
Izmailov A.G. ― Cand. Med. Sc., Associate Professor of the Department of General Surgery, tel. +7-927-412-73-46, e-mail: izmailov_alex@mail.ru
Dobrokvashin S.V. ― D. Med. Sc., Professor, Head of the Department of General Surgery, tel. (843) 557-39-46, e-mail: gsurgery1@yandex.ru
Volkov D.E. ― Cand. Med. Sc., Associate Professor of the Department of General Surgery, tel. (843) 556-29-26, e-mail: gsurgery1@yandex.ru
Lukoyanychev E.E. ― Cand. Med. Sc., surgeon, tel. +7-910-79-40-549, e-mail: egor-lukoyanychev@yandex.ru
Pyrkov V.A. ― Head of the Surgical Department, tel. (843) 557-39-26, e-mail: pyrkovgvv@rambler.ru
Zakirov R.F. ― Cand. Med. Sc., oncologist, doctor of the Surgical Department, tel. (843) 557-39-26, e-mail: zakmurat@yandex.ru
Davlet-Kildeev Sh.A. ― doctor of Surgical Department, tel. (843) 556-29-26, e-mail: i.fydfhsx51@yadex.ru
The article presents the results of applying the newly developed device for punctures of postoperative cavities in patients with ventral hernias. The device includes a needle and a 15 ml container with antiseptic solution, which form a single sealed complex, additionally equipped with a pushing pad, a support pad and a handle; the container is formed as a needle-scalpel with the longitudinal bore and an aperture at the base, which is rigidly fastened to the pushing pad. The device was used in the postoperative period in 3 (9.4%) patients of the main group with the purpose of emptying the cavity structures. In contrast to the traditional method of evacuation, it took 1-2 punctures to resolve the postoperative wound.
Key words: ventral hernia, seroma, device for punctures.
REFERENCES
- Abbaszade T.N., Andreev A.I., Anisimov A.Yu. Prevention of early postoperative wound complications when large hernia repair ventral hernias. Vestnik sovremennoy klinicheckoy meditsiny, 2015, vol. 8 (pril. 1), pp. 11-18 (in Russ.).
- Fatkhutdinov I.M., Krasil’nikov D.M. The choice of hernia repair in patients with strangulated extensive and giant postoperative hernias. Sovremennye problemy nauki i obrazovaniya, 2008, no. 2, pp. 41-43 (in Russ.).
- Lavreshin P.M., Efimov A.V., Gobedzhishvili V.K. et al. Prevention of postoperative wound complications in patients with postoperative ventral hernias. Vestnik natsional’nogo mediko-khirurgicheskogo tsentra im. N.I. Pirogova, 2015, no. 2, pp. 61-64 (in Russ.).
- Mukhtarov Z.M., Malkov I.S., Alishev O.T. Prevention of inflammatory wound complications after surgery with postoperative ventral hernias. Prakticheskaya meditsina, 2014, no. 5 (81), pp. 106-109 (in Russ.).
- Izmaylov A.G., Dobrokvashin S.V., Volkov D.E., Lukoyanychev E.E. Profilaktika gnoyno-vospalitel’nykh ranevykh oslozhneniy posle operativnykh vmeshatel’stv pri posleoperatsionnykh ventral’nykh gryzhakh. Materialy IX Vserossiyskoy konferentsii obshchikh khirurgov s mezhdunarodnym uchastiem. Yaroslavl, 2016. Pp. 686-688.
- Chevrel J.P., Bendavid R. et al. (eds). Treatment of incisional by an overlapping herniorrhaphy and onlay prosthetic implant. Abdominal wall hernias: preinciples and management. New York: Springes-Verlag, 2001. Pp. 500-503.
- Lukoyanychev E.E., Izmaylov S.G., Izmaylov A.G., Sinyagina V.G. Patent na izobreteniya RF №2526246. Ustroystvo dlya vskrytiya polostnykh obrazovaniy. 26 iyunya. 2014 g [The patent for the invention of the Russian Federation no. 2526246. A device for opening the abdominal structures. June 26, 2014].